3 Vacuum system

3.3 Disinfection with a wet suction system

3.3
Disinfection with a wet suction system
8
simultaneously
Care and cleaning instructions for the practice team C2
After the cleaning process, you must fill at least
1 liter of water into the cuspidor and aspirate it in or-
der to remove any disinfectant residues from the
suction hoses.
At the end of the cleaning process, pull off the hoses
and put them back into their holders.
Remove the cleaning adapter.
Finally, reinsert the gold trap.
With this system, the automatic separator and, where
appropriate, the amalgam separator are not integrated in
the water unit. Both separation processes are performed
centrally.
Daily during the noon break and at the end of
each workday
Flush the vacuum system with a suitable cleaning
and disinfecting agent as described under 3.2 Disin-
fection with a dry suction system.
After the liquid has been aspirated through the two
suction hoses, remove the cleaning adapter.
Then pour 1 liter of disinfectant solution into the cus-
pidor while simultaneously pressing the button (ar-
row) for approx. 10 seconds.
Pressing the button enables the cleaning and disin-
fectant liquid from the cuspidor to reach the vacuum
line via the cuspidor valve.
Finally, reinsert the gold trap.
ATTENTION
Use only cleaning and disinfecting agents which have
been approved by Sirona (see page 2, General informa-
tion)!
Do not use any washing powder or other disinfec-
tants!
Sirona Dental Systems GmbH
+
+
, M1
59 58 512 D 3370
D 3370.103.01.13.02