Amalgam Separation; Replacing The Amalgam Separator - Sirona C2+ Cleaning And Care

Hide thumbs Also See for C2+:
Table of Contents

Advertisement

4 Amalgam separation

4.1 Replacing the amalgam separator

4
Amalgam separation
4.1
Replacing the amalgam separator
1
Amalgam
2
Grease
7
12
A B
AMALG
3
2
4
!
O-ring
3
Use Sirona
Replacement amalgam
Order No.: 14 34 138
only!
Disposal
Care and cleaning instructions for the practice team C2
When AMALG lights up on the display on the dentist ele-
DESINF
ment, the amalgam separator must be replaced.
RPM
To replace the amalgam separator, proceed as follows:
Disinfect as described in Chapter 3 "Vacuum sys-
tem" .
i
NOTE
Always wear work gloves when performing these tasks!
Open flap (1) at the base of the water unit.
Do NOT turn the unit off!
Turn the lower part (2) of the amalgam separator
with rotor (3) in the direction of the arrow until it can
be removed easily.
AMALG starts flashing on the dentist element
display and an audible signal sounds.
Put any amalgam residues from the suction hoses
and the cuspidor into the dismantled rotor for dispos-
al (see pages 6 and 10).
Hold the bottom part upright and click on the
transport cap (4). This releases a bonding agent.
Therefore, do not remove the transport cap
again!
Remove the rotor with the cap from the lower part
and place it in the special packaging for disposal
Insert the new rotor (3) in the lower part (2).
Do not install a used or recycled rotor!
Reinstall the lower part and turn it in the opposite di-
rection of the arrow up to the stop. The bar (7)
should point to the chair at the start and to the siphon
(6) after it is screwed tight.
If, after the rotor is inserted, AMALG still flashes
on the display and the audible signal continues,
the lower part is not correctly locked into place.
Close the flap (1) again.
Sirona Dental Systems GmbH
+
+
, M1
.
59 58 512 D 3370
D 3370.103.01.13.02