Sirona Dental Systems GmbH
Care and cleaning instructions for the practice team C2
1
Connection box
59 58 512 D 3370
D 3370.103.01.13.02
+
+
, M1
Replacing the water and air filters
If you notice any changes in media flows, please check
the water and air filters (1) for permeability. Replace
them if necessary.
•
Switch the main switch OFF.
•
Pry off cover 2 with a screwdriver.
•
Remove screw caps 3 and replace filters 1.
OFF
For reorders:
Filters:
Main fuse
The main fuse (4) is located below the air filter.
For reorders:
Unit main fuse for
230VAC:
Unit main fuse for
100VAC/115VAC:
1 Connection box
Order No. 14 43 436
T 6.3AH, 250VAC,
Order No. 10 77 452
T 10AH, 250VAC,
Order No. 10 77 460
5