1529 Chub-E4 Thermometer Readout
User's Guide
7.4.2.2
62
The LOG PER parameter is for selecting the interval between measurements.
You may select between 0.1, 0.2, 0.5, 1, 2, 5, 10, and 30 seconds, 1, 2, 5, 10,
and 30 minutes or 1 hour. The LOG PER must be set to a value less than or
equal to the measurement period. The 1529 will not log readings faster than the
measurement period. The measurement period is set in the MEASURE PE-
RIOD function of the CHANNEL menu and is the rate at which readings are
taken.
If the log period is set to a value less than the measure period the following
message is displayed in the lower portion of the display.
DATA WILL BE LOGGED NO FASTER THAN
THE MEASURE PERIOD (X SEC ).
Press ENTER to continue...
The X in the message is the current measurement period. After the Enter but-
ton is pressed, the measurements are logged at the current measure period.
The COUNT parameter is for selecting the total number of readings to log. The
default is 8,160, which is the maximum number of readings that can be logged.
If the count is set to 10 and two channels are displayed on the screen at the time
logging starts, five readings from each channel are logged for a total of 10 read-
ings. The auto log automatically stops after the number of readings selected is
reached or when the user stops the log manually.
Press the LR buttons to change a setting and the UD buttons to scroll be-
tween the parameters. Press Enter button to save the new setting. Press and
hold the Exit button to cancel to the main display or press the EXIT button to
cancel and to move to the next parameter.

Start Stop

The START STOP function starts and stops auto log sessions. When this func-
tion is selected, the user is prompted to either start or stop an auto log session
by pressing the Enter button. If an auto log session is not currently running,
the user is prompted to START RECORDING DATA. If an auto log session is
running, the user is prompted to STOP RECORDING DATA.