Bump Test; Settings; Amplitude Trigger Setting - Adash VA4 Pro II User Manual

Vibration analyzer
Table of Contents

Advertisement

ADASH

Bump Test

It enables to find the resonance frequency of machine body. Set the used channel and the number of hits for
averaging.
Test contains two parts. The level of amplitude trigger is found in the first part. The response spectrum is
developed in the second part.

Settings

Channel
Hits

Amplitude trigger setting

In the first part of test we need to detect the correct value for triggering. The system set it automatically. Make
the hit and look the moving of the hand on display. Make all hits with the same force. More then one hit is
recommended (3-4). Push Stop to finish it.
For further technical and contact information visit
channel with the accelerometer
number of hits
124
www.adash.com
or email: info@adash.com