Ac Inlet Assembly - Respironics BiPAP Vision Service Manual

Hide thumbs Also See for BiPAP Vision:
Table of Contents

Advertisement

5.25 AC Inlet
Assembly
1062788
Follow these steps to remove the AC inlet assembly (Figure 5-33). Reverse to
install.
1. Remove the blower (section 5.12).
2. Remove the screws and nuts that attach the AC inlet and power line
filter to the back panel.
3. Disconnect the transformer cable.
4. Remove the nut that attaches the ground wires from the bottom
enclosure.
4-40 x 3/8 screw (x2)
(6 lbf-in.)
Transformer cable connection
Power line filter
4-40 locknut (x2)
(6 lbf-in.)
AC inlet
Ground wires (from AC inlet and PSS)
#8 washer (x1)
Figure 5-33: Removing the AC Inlet Assembly
BiPAP Vision Service Manual © Respironics, Inc.
Component Removal/Installation
6-32 locknut (x2)
(10 lbf-in.)
(Note wire position)
8-32 nut (x1 per ground wire)
(10 lbf-in.)
Chapter 5
5-35

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents