Replacing The Inlet Filter - Respironics BiPAP Vision Service Manual

Ventilatory support system
Hide thumbs Also See for BiPAP Vision:
Table of Contents

Advertisement

5.2 Replacing the Inlet Filter

CAUTION:
A dirty inlet filter may cause high operating temperatures, and may affect ventilator
performance. Examine the inlet filter for integrity and cleanliness before each use, and
as required during operation.
Step 1
Turn the BiPAP Vision OFF and unplug the electrical cord from the wall outlet and from the
back of the unit.
Removing the Filter
Step 2
Remove the inlet filter cap by pinching the latch, then rotate the cap until the hinge is
free from its slot.
NOTE:
The inlet filter is disposable. Do not attempt to clean the inlet filter. When the filter is
dirty replace it with a new filter. Use only Respironics filters; see Chapter 7 for
the filter reorder number.
(P/N 582101 x6)
Installing the Filter
Step 3
Place the filter inside the cap, then reverse Step 2 to reinstall the filter cap.
NOTE:
To clean any of the accessories, refer to each accessory's instruction sheet.
Figure 5-1
Removing the Filter Cap
(P/N 1003444)
BiPAP Vision Service Manual
Chapter 5: Routine Maintenance
5-3
1015010