Chapter 4. Periodic Maintenance

4.1 Cleaning

1062788
Table 4-1 summarizes BiPAP Vision periodic maintenance procedures.
Table 4-1: Periodic Maintenance
Frequency
Component
As required
BiPAP Vision
ventilator
Power cord
Air inlet filter
element
Nylon mesh inlet
filter
Oxygen regulator
filter
Fuses
Internal alarm
battery
Annually
Audible alarm
Blower
BiPAP Vision
ventilator
Record maintenance activities on the periodic maintenance log (section 4.7).
Before cleaning, turn the ventilator off and disconnect the power cord from the
back of the unit and the wall outlet.
Front panel
Clean as needed by wiping with water or 70% isopropyl alcohol.
Enclosure
Clean exterior as needed by wiping with any antibacterial agent. Clean the
Auto-Trak sticker with mild soap and water.
Interior
Carefully clean accessible interior areas with and ESD-safe vacuum
cleaner.
CAUTION:
Do not immerse the device in water or allow any liquid to enter the
enclosure or inlet filter. These instructions are for the BiPAP Vision
ventilator only. To clean accessories, see the applicable instructions.
BiPAP Vision Service Manual © Respironics, Inc.
Maintenance
Clean exterior surfaces (section 4.1).
Inspect and replace if damage or wear is visible.
Replace (section 4.2).
Clean/replace (section 4.3).
Replace (section 4.4).
Replace (section 4.5).
Recharge (section 4.6)
Activate Test Alarms (Chapter 7) to verify audible
and visual function.
Perform blower valve calibration (Chapter 7).
Clean ventilator interior and exterior as required
(section 4.1). Complete system final test (Chapter
7).
4-1