Audible Alarm Current Check; Safe State Power On; Watchdog And Low Voltage Reset - Respironics BiPAP Vision Service Manual

Hide thumbs Also See for BiPAP Vision:
Table of Contents

Advertisement

1062788
3.4.19

Audible Alarm Current Check

An internal test verifies that the audible alarm current is acceptable.
3.4.20

Safe State Power On

The DC board contains circuitry that power on the hardware in a safe state. In
a safe state, the backlight is off, the display is off, and the ICB is terminated.
When the MC board determines that no Vent Inop error exists, it allows the
device to resume operation under normal operating conditions.
3.4.21

Watchdog and Low Voltage Reset

The watchdog function must be periodically reset by the microprocessor if a
time-out period has been exceeded. This function resets the processor if the
software stops functioning correctly. When a low logic level is detected, the
ELC is activated and the system shuts down.
BiPAP Vision Service Manual © Respironics, Inc.
Chapter 3
Theory of Operation
3-17