Troubleshooting Vent Inop Error Codes - Respironics BiPAP Vision Service Manual

Hide thumbs Also See for BiPAP Vision:
Table of Contents

Advertisement

6.5 Troubleshooting
Vent Inop Error Codes
Vent Inop indicator lights,
audible alarm sounds.
Verify that input voltage
is connected.
ventilator power on
Select Monitoring > Options > Error
Messages: the error at the top of the
list is the Check Vent code.
See troubleshooting information on
Vent Inop error codes (Table 6-4).
Remove top enclosure and ensure
that all connections are secure and
subsystems are correctly installed.
Perform corrective action.
Verify that problem is corrected.
Replace top enclosure.
Perform applicable testing and
calibration (Chapter 7).
1062788
This flow chart (Figure 6-3) summarizes the troubleshooting sequence to
follow if the Vent Inop indicator lights.
Does
No
and display
properly?
Yes
Figure 6-3: Vent Inop Troubleshooting Flow Chart
BiPAP Vision Service Manual © Respironics, Inc.
Troubleshooting
Remove top enclosure and ensure that
all connections are secure and
subsystems are correctly installed.
Connect service laptop to the
subsystem whose error LED is lit
(Chapter 7 describes setup).
See troubleshooting information on
Vent Inop error codes (Table 6-4).
Chapter 6
6-13