5.6 Top Enclosure

Top enclosure
Screw (x2)
Bottom enclosure
1062788
Follow these steps to remove the top enclosure (Figure 5-10). Reverse to
install.
NOTE:
Apply a small amount of threadlock reinstalling the top enclosure screws
to the back panel.
1. Turn the ventilator off and disconnect the power cord from the back of
the device.
2. Remove the 2 screws that attach the top enclosure to the bottom
enclosure.
3. Slide the enclosure straight back, then lift slightly to clear oxygen
module screws.
4. Pull the enclosure back to disengage the latch that attaches the top
enclosure to the bottom enclosure, then lift the top enclosure away
from the device.
Figure 5-10: Removing the Top Enclosure
BiPAP Vision Service Manual © Respironics, Inc.
Component Removal/Installation
Chapter 5
5-11