Record DC board error codes, if any:
Record software version:
Oxygen module installed?
Alarm module installed?
FREEZE/UNFREEZE functions correctly
SCALE functions correctly, waveform display changes
Alarms Test
Alarm occurs upon tubing disconnect, alarm message displayed
After reconnecting tubing Alarm reset clears alarm
Alarm sounds and wrench icon lights when AC cord removed
When powered up with AC cord installed, no audible alarm
Oxygen Module (OM) Test
%O
setting
2
21%
30%
O
cylinder valve closed, O
2
40%
60%
80%
100%
PAV/T Mode Test
PAV/T mode features displayed correctly, no alarms occur
Earth Resistance and Leakage Current Test
Test
Earth resistance
Normal pole, no earth, L2
Reverse pole, no earth, L2
Reverse pole, no earth, no L2
Normal pole, no earth, no L2
Nurse Call/Remote Alarm Test
Correct continuity with ventilator off and with active alarm
Tested by:
(Print name)
1062788
O
analyzer reading
2
flow alarm active within 15 breaths
2
DMM reading
(Signature)
BiPAP Vision Service Manual © Respironics, Inc.
Yes / No
Yes / No
Pass
Fail
Pass
Fail
Pass
Fail
Pass
Fail
Pass
Fail
Pass
Fail
Specification
Pass/Fail
18.0 - 24%
Pass
Fail
27.0 - 33%
Pass
Fail
Pass
Fail
36 - 44%
Pass
Fail
54 - 66%
Pass
Fail
72 - 88%
Pass
Fail
> 90%
Pass
Fail
Pass
Fail
Specification
< 0.20 Ohms
< 300 uA
< 300 uA
< 1000 uA
< 1000 uA
Pass
Fail
Testing and Calibration
N/A
Pass/Fail
Pass
Fail
Pass
Fail
Date:
Chapter 7
7-35