Chapter 7
Testing and Calibration
Ventilator serial no.
Model no.
Power Indicator and LCD Controls Test
Power indicator lights
Brightness control function
Contrast control function
Pressure Accuracy Test
CPAP setting
Set
5 cmH
O
2
10 cmH
O
2
20 cmH
O
2
Flow Accuracy Test
Flow setting
Flowmeter
0 LPM
10 LPM
60 LPM
120 LPM
120 LPM
Dynamic Pressure Regulation Test
Exhalation port test complete
Flow waveforms created (peak flow ~ 100 LPM, rate ~ 30 BPM)
Manometer reading = 18-22 cmH
S/T Mode Performance Test
Vent cycles between low and high pressure
IPAP = 33-37 cmH
O, EPAP = 3-7 cmH
2
Rate = 9.3-10.8 BPM
Rate reading:
Options and Controls
Invert, Bar Graph, and Test Alarms function correctly
Record displayed error messages, if any:
Record MC board error codes, if any:
Record PC board error codes, if any:
7-34
BiPAP Vision System Final Test Form
Pass
Fail
Pass
Fail
Pass
Fail
Outlet
Unit
Compensated total flow
O
2
Pass
O
Pass
2
Pass
BiPAP Vision Service Manual © Respironics, Inc.
Service notification
(Respironics use only)
Total operating time
Patient
Manometer
Specification
3-7 cmH
8-10 cmH
18-22 cmH
Pressure
Specification
-5 to 5 LPM
4.2-15.8 LPM
50.2-69.8 LPM
105.6-134.4 LPM
18-22 cmH
Pass
Fail
Pass
Fail
Pass
Fail
Fail
Fail
Fail
Pass
Fail
Pass
Fail
Pass/Fail
O
Pass
Fail
2
O
Pass
Fail
2
O
Pass
Fail
2
Pass/Fail
Pass
Fail
Pass
Fail
Pass
Fail
Pass
Fail
O
Pass
Fail
2
1062788