5.23 RS-232

Connector

Back panel (outside)
1062788
Follow these steps to remove the RS-232 connector (Figure 5-31). Reverse to
install.
NOTE:
This procedure applies only to units built s/n 106000 and above (units
built s/n 105999 and below do not include an RS-232 connector).
1. Remove the blower (section 5.12).
2. Remove the RS-232 cable from MC J3.
3. Remove the 2 screws that attach the RS-232 connector to the outside
of the back panel.
4. Feed the connector and cable through the cutout in the back panel.
RS-232 connector
M2.5 x 8 screw (x2)
Figure 5-31: Removing the RS-232 Connector
BiPAP Vision Service Manual © Respironics, Inc.
Component Removal/Installation
RS-232 connector (red
cable stripe points up)
Cable connects
to MC J3
Chapter 5
Back panel
(inside)
5-33