Chapter 3
Theory of Operation
3-24
Figure 3-10: MC Board Jumper Locations
CAUTION:
The BiPAP Vision nurse call/remote alarm port must be connected to
nurse call systems that meet relevant local safety standards.
The nurse call port must be connected to a low voltage circuit
(< 42.4 Vpeak AC or 50 VDC). Leakage currents from the low voltage
circuit must not cause the device leakage current to exceed acceptable
levels. The rated output current of the low voltage circuit must not exceed
1 A.
BiPAP Vision Service Manual © Respironics, Inc.
JP2
JP1
1062788