Chapter 5
Component Removal/Installation
5.24 Nurse Call/
Remote Alarm
Connector
Back panel
(outside)
5-34
Follow these steps to remove the nurse call/remote alarm connector
(Figure 5-32). Reverse to install.
NOTE:
This procedure applies only to units built s/n 106000 and above (units
built s/n 105999 and below do not include a nurse call/remote alarm
connector).
1. Remove the blower (section 5.12).
2. Remove the nurse call/remote alarm cable from MC J5.
3. Remove the 9/16-in. nut that attaches the nurse call/remote alarm
connector to the inside of the back panel.
4. Feed the connector and cable through the cutout in the back panel.
Nurse call/remote
alarm connector
Cable connects
to MC J5
Figure 5-32: Removing the Nurse Call/Remote Alarm Connector
BiPAP Vision Service Manual © Respironics, Inc.
Nut (x1)
(10 lbf-in.)
Back panel
(inside)
1062788