Chapter 7
Testing and Calibration
7-32
7.7.12

Nurse Call/Remote Alarm Test (if applicable)

NOTE:
This test applies only to ventilator s/n 106000 and above. This optional
test provides a way to verify the nurse call/remote alarm connector as
required by institutional protocol.
This test allows testing for the nurse call/remote alarm connector when
configured as:
Remote alarm (option 1)
Nurse call, normally closed (option 2)
Nurse call, normally open (option 3), the factory-set configuration.
Jumpers on the MC board configure the nurse call/remote alarm connector
(Figure 7-10):
Nurse call/remote alarm
configuration
Remote alarm (option 1)
Nurse call, NC (option 2)
Nurse call, NO (option 3)
NOTE:
.
Figure 7-10: MC Board Jumpers
BiPAP Vision Service Manual © Respironics, Inc.
JP1
2,3
2,3
1,2
Before testing, ensure that MC
board jumpers are set correctly.
JP2
JP1
JP2
2,3
1,2
1,2
1062788