4.7 Periodic
Maintenance Log
1062788
Use this log to record maintenance activities. Service intervals may vary
according to institutional protocol, and comply with any applicable regulations
that deviate from this schedule.
Model no. ______________________ Serial no. _______________________
Maintenance activity
BiPAP Vision ventilator: clean exterior.
Power cord: inspect for damage.
Inlet filter: replace.
Nylon mesh air inlet filter: clean/re-
place.
Oxygen regulator filter: replace.
Audible alarms: activate Test Alarms to
verify.
Blower: perform blower valve calibra-
tion.
BiPAP Vision ventilator:
Clean exterior.
Vacuum dust/lint from
interior.
Complete system final test.
Record hours of operation (displayed
on Options screen)
Tested by: ________________________________ Date: _______________