Chapter 2. Warnings, Cautions, and Notes

2.1 Warnings

1062788
Throughout this manual the following definitions apply:
WARNING:
A condition that could cause injury to a patient or operator if the operating
instructions in this manual are not followed correctly.
CAUTION:
A condition that could cause damage to, or shorten the service life of, the
device.
NOTE:
Important information concerning the device.
Do not use the BiPAP Vision in the presence of a flammable
anesthetic mixture with air, oxygen, or nitrous oxide.
Oxygen supports combustion. Do not use oxygen while smoking or
in the presence of an open flame.
The BiPAP Vision does not provide an oxygen sensor to monitor
oxygen concentrations delivered to the patient circuit. Use
oximetry to monitor the use of oxygen with the BiPAP Vision.
If the Vent Inop (wrench) icon lights, see Chapter 6 of this manual
for troubleshooting information.
Never attach oxygen tubing or any positive pressure source to the
pressure port on the front panel of the BiPAP Vision.
Do not attempt to use the RS-232 connector on the back panel of
the device to obtain repair information during operation on a
patient.
To ensure personal safety and correct device performance, only
qualified technicians are to perform repairs to the BiPAP Vision.
Contact Respironics for service training and authorization
information.
High voltages are present inside this device. To avoid electrical
shock, disconnect the electrical supply before attempting any
repairs on the device.
For continued protection against risk of fire, replace fuses with
those of the same type and rating only.
To avoid electric shock, unplug the BiPAP Vision before cleaning.
BiPAP Vision Service Manual © Respironics, Inc.
2-1