6.1.5 R
EPLACING THE
Kit
PCA
T
PCA:
O REMOVE TH
1. Remove all components as instructed in the previous sections.
2. Remove the two # 4 x 1/2" screws that secure the PCA to the Blower Cap using a T-10 Torx driver.
3. Lift the PCA up away from the device.
4. Disconnect the DC Connector and Blower Connector from the PCA.
DC Connector
PCA
Included in Kit
PCA
# 4 x 1/2" screws (qty 2)
Blower Connector
F
IGURE
Torque screws to 2.5 IN-LB
when installing PCA.
6-8: PCA
Tools Required
Torx Screwdriver (T10 is
recommended)
P
6-9