Respironics DreamStation Service & Technical Reference Manual page 63

Table of Contents

Advertisement

P
5-18
AGE
P
/F
ROBLEM
AILURE
General Noise
Whistling Noise from Humidifier
Whistling Noise from PAP Device
M
P
ODE
OSSIBLE
Humidifier Seals are
damaged or not seated
Humidifier Back Panel not
secure
Enclosures not secure
Blower Box is cracked
Blower Isolators not
seated
Blower is contaminated
Blower noisy in general
Something fell inside
Blower Box
Too much Blower wire
slack outside of the
Blower Box (pulled too
tight - causing vibration)
Humidifier Seals are
damaged or not seated
Humidifier Back Panel not
seated
Wire guard loose
Tank is cracked
Dry Box not seated
Flip Lid latch damaged or
not latching well
PAP and Humidifier not
fully connected
PAP and Humidifier not
fully connected
PAP Enclosure
separated.
Blower noisy
C
AUSE
Make sure all Seals are
seated.
Make sure PCA is aligned on
seals.
Make sure enclosures are
secure.
Check for any component
damage.
Check for loose components.
Check Blower wire length
outside of Blower Box (should
measure approx. 9.5 inches).
Replace any failed/damaged
component.
Make sure all Seals and
components are seated.
Check for any component
damage.
Replace damaged
components.
Check all connections.
Replace blower if noisy.
1117539, V
S
T
TEPS TO
AKE
. 07
ER

Hide quick links:

Advertisement

Table of Contents
loading