LERT
Incorrect
Power
Supply
Low Voltage
Automatic Off
Inlet blocked.
Check filter.
Low Leak:
Check Mask
and Tube
I
T
CON
YPE
Notification
Notification
Status
Notification
Notification
D
ESCRIPTION
Indicates an
incompatible power
supply is attached.
Low voltage.
Displayed when
therapy ends due to
automatic off
function.
Blocked airway
Blocked airway
P
OSSIBLE
A
C
AUSE
Incompatible
Confirm power
power supply,
cord is full
or power cord is
inserted into
not fully inserted
device's power
into device's
inlet. Confirm a
power inlet.
compatible
Philips
Respironics
power supply is
attached.
Switch to
compatible
power supply if
needed.
Incompatible
Confirm a
power supply is
compatible
attached.
Philips
Respironics
power supply is
attached.
Switch to
compatible
power supply if
needed. If
battery is being
used, ensure
battery is
adequately
charged.
The mask has
n/a
been removed.
Blockage at
Check device air
device
inlet is not
inlet.
obstructed.
Check air
filter(s) are
installed properly
and are clean;
replace if
needed.
Blockage at
n/a
tube or mask.
P