Final testing of a device is a mandatory requirement after any repairs are made to the device, or if the device
enclosure has been opened for any reason. Please note that the firmware is updated using the below test
system if it is not currently up-to-date.
8.1 R
E
EQUIRED
Hardware:
DreamStation Power Supply, 80W, (PN 1118499), Qty 1 ea.
DreamStation Serial Accessory Module (PN 1123881 or PN 1126240 (includes serial and micro-
USB cables listed below)), Qty 1 ea.
Serial RS232 Cable DB9F-DB9M, 6 ft (PN 1037268), Qty 1 ea.
USB Cable, Type A to Micro B (PN 1104765), Qty. 1 ea.
Hose 22mm-18 in. (PN 1008198), Qty 1 ea.
15mm Heated Tube (PN HT15), Qty 1 ea.
German Leak Device (ISO Port) (PN 1127131), Qty 1 ea.
Pollen Filter (PN 1122446), Qty 1 ea.
SD Card (PN 1063859 - 10 pk), Qty 1 ea.
Printer (network or local connection)
PC with the following minimum requirements:
Windows 7 32/64 Bit Operating System, Professional (Enterprise can be used for Philips net-
work users only)
3.20 GHz Processor
100 GB Free hard drive space
8 GB RAM
1 serial port
Minimum 4 USB port
Keyboard, mouse and 17 inch monitor
Supports 220 VAC +/- 10% and 50 Hz +/- 1Hz, or 120 VAC +/- 10% and 60 Hz +/- 1Hz power
Software (download and install from my.respironics.com):
NI LabVIEW 2014 Support Package (located under the Utility Tools category)
DreamStation Device Driver Package, Version 1.0 or higher (located under the DreamStation Ser-
vice Tools and Documents category)
DreamStation Field Service Application (FSA), Version 1.0 or higher (located under the DreamSta-
tion Service Tools and Documents category)
ESTING AND
Testing must occur in a lab equipment environment only.
QUIPMENT