1117539, V
. 07
ER
5.6 D
A
EVICE
LERTS
Device alerts are pop-ups that show up on the UI screen. There are 5 types of alerts described here:
•
Status: These alerts are just the pop-up screen.
•
Notification: These alerts consist of the pop-up screen in addition to a blinking Power LED on top
of the device.
•
Alert 1: These alerts consist of the pop-up screen, a blinking Power LED and an audible beep
when displayed. This alert will not occur during therapy.
•
Alert 2: These alerts consist of the pop-up screen, a blinking Power LED and an audible beep
when displayed. This alert can occur during therapy.
•
Safe State: These alerts consist of the pop-up screen, a blinking Power LED and a repeating audi-
ble beep. Note: Status alerts automatically time out after 30 seconds and their pop up screens dis-
appear. All other alerts must be acknowledged to clear.
Alert Summary Table: The following table summarizes the alerts.
A
LERT
Data Activity:
Do not
remove SD
card.
Change
Accepted
EZ-Start
Pressure
Incremented
to xx.x
I
T
CON
YPE
Status
Status
Status
D
ESCRIPTION
SD
card
read/write
underway.
Confirms acceptance
of prescription change
or device upgrade.
Displays when
EZ-Start mode is
enabled and
device is
increasing therapy
pressure setting
for the next
session.
P
OSSIBLE
A
C
AUSE
n/a
No action
needed.
n/a
No action
needed.
No action needed.
n/a
P
5-23
AGE
CTION