P
5-2
AGE
5.2 V
P
ERIFYING
•
If the device fails to perform within the stated specifications, have the
system serviced by a qualified Philips Respironics-approved service
facility.
•
You will need the following equipment to verify the pressure:
•
Philips Respironics Pressure Calibration Kit
Kit Includes:
•
Philips Respironics Whisper Swivel II (1)
•
Philips Respironics O2 Enrichment Final Assembly (2)
•
Closed end cap (3)
•
Philips Respironics flexible tubing (4)
•
Pressure tubing (5)
•
Philips Respironics Digital Manometer (6) or equivalent
Minimum Specifications:
0 - 25 cm H2O (or better)
±0.3 cm H2O accuracy
±0.1 cm H2O resolution
•
Blue pollen filter (not shown)
RESSURE
F
5-1: P
IGURE
WARNING
V
RESSURE
ERIFICATION
1117539, V
. 07
ER