Verifying Pressure - Respironics DreamStation Service & Technical Reference Manual

Table of Contents

Advertisement

P
5-2
AGE
5.2 V
P
ERIFYING
If the device fails to perform within the stated specifications, have the
system serviced by a qualified Philips Respironics-approved service
facility.
You will need the following equipment to verify the pressure:
Philips Respironics Pressure Calibration Kit
Kit Includes:
Philips Respironics Whisper Swivel II (1)
Philips Respironics O2 Enrichment Final Assembly (2)
Closed end cap (3)
Philips Respironics flexible tubing (4)
Pressure tubing (5)
Philips Respironics Digital Manometer (6) or equivalent
Minimum Specifications:
0 - 25 cm H2O (or better)
±0.3 cm H2O accuracy
±0.1 cm H2O resolution
Blue pollen filter (not shown)
RESSURE
F
5-1: P
IGURE
WARNING
V
RESSURE
ERIFICATION
1117539, V
. 07