Respironics DreamStation Service & Technical Reference Manual page 58

Table of Contents

Advertisement

1117539, V
. 07
ER
C
E
ODE
RROR
E-85
ERR_FLOW_SENSOR_O
CCLUDED
E-87
ERR_FLOW_SENSOR_B
US
E-93
ERR_RTC_VALUE
E-94
ERR_RTC_STOPPED
E-100
ERR_HUMID_NO_HEAT
E-101
ERR_HUMID_MAX_TEMP
E-105
ERR_HUMID_AMBIENT_
COMM
E-106
ERR_HEATED_TUBE_MA
X_TEMP
E-107
ERR_HEATEDTUBE_DIS
CONNECT
E-108
ERR_HUMIDIFIER_DISC
ONNECT
E-130
ERR_TASK_WDOG_TIME
OUT
E-131
ERR_WIN_WDOG_TIME
OUT
E
RROR
N
AME
L
EVEL
CONTINUE
CONTINUE
CONTINUE
CONTINUE
CONTINUE
CONTINUE
CONTINUE
CONTINUE
CONTINUE
CONTINUE
REBOOT
ABORT
F
AILED
C
OMPONENT
PCA
1.Clear error log and test.
2. If the test fails or the error
reoccurs, replace the PCA
and retest.
PCA
1. Clear error log and test.
2. If the test fails or the error
reoccurs, replace the PCA
and retest.
N/A
1. Clear error log, set the
RTC and test.
2. If the test fails or the error
reoccurs, replace the PCA
and retest.
PCA
1. Replace PCA and test.
Heater Plate
1. Clear error log and test.
2. Test humidifier separately.
Heater Plate
1. Clear error log and test.
2. Test humidifier separately.
PCA
1. Clear error log and test.
2. If the test fails or the error
reoccurs, replace the PCA
and retest.
Heated Tube
1. Clear error log and test.
2. Test humidifier separately.
N/A
1. Clear error log and test.
N/A
1. Clear error log and test.
PCA
1. If error was NOT Last
Stop Error, clear error log
and test.
2. If error was Last Stop
Error, replace PCA and test.
PCA
1. If error was NOT Last
Stop Error, clear error log
and test.
2. If error was Last Stop
Error, replace PCA and test.
P
5-13
AGE
A
CTIONS