Respironics DreamStation Service & Technical Reference Manual page 65

Table of Contents

Advertisement

P
5-20
AGE
P
/F
ROBLEM
AILURE
The Display is Erratic
Back Light Does Not Work
Back Light Dims Too Quickly
Back Light Does Not Dim
Can't Change Humidity Settings
M
P
ODE
OSSIBLE
The device has been
dropped or mishandled
The device is in an area
with high Electromagnetic
Interference (EMI)
emissions.
Problem with UI ribbon
cables and connections.
Faulty PCA
Ambient light sensor is
blocked.
User perception
Ambient light sensor not
functioning
Ambient light sensor not
functioning
User perception - can
only change setting when
therapy is active. You
cannot change the
settings in the device
menus
Dial does not function
C
AUSE
Unplug the device and reapply
power.
Relocate the device to an area
with lower EMI emissions
(cellular phones, computers
TVs, etc.).
Check UI connections.
Replace UI Panel if faulty.
Replace PCA if ambient light
sensor is faulty.
Verify nothing is blocking the
ambient light sensor.
Verify ambient light sensor is
functioning by covering the
sensor hole on the Top
Enclosure.
Replace PCA if ambient light
sensor does not function.
Verify ambient light sensor is
functioning by covering the
sensor hole on the Top
Enclosure.
Replace PCA if ambient light
sensor does not function.
Activate therapy and verify the
settings change.
Verify the dial functions.
Replace any damaged UI
panel if dial does not function.
1117539, V
S
T
TEPS TO
AKE
. 07
ER