Disinfection; After Disinfection - Respironics DreamStation Service & Technical Reference Manual

Table of Contents

Advertisement

1117539, V
. 07
ER
7.5.2 D
ISINFECTION
The recommended disinfection methods are identified below.
Thermal Disinfection: Immersion in a (tap) water bath at 75° C ± 2° C for 30 minutes
Cidex
Cidex OPA
Follow all instructions from the manufacturer of the treatment products. Any deviation from
these instructions, the manufacturer's instructions, or agents not listed in this guide may
impact the performance of the product. Review all applicable instructions for additional
warnings and cautions.A brush is not necessary for the inside of the tube of the heated
tubing.
Treat the tank, seal and heated tubing as different medical devices when rinsing them with
water. For example, the tank, seal and heated tubing cannot be rinsed in the same volume
of water.
Use 3 gallons (11.4 liters) as the rinse volume when following the instructions.
7.5.3 A
D
FTER
ISINFECTION
1. Inspect the tank, seal and heated tubing for damage or wear (cracking, crazing, tears, or damage,
etc.). Discard and replace if damaged.
2. After the final water rinse described in the chemical disinfection instructions, separately rinse the
tank, seal and heated tubing a minimum of 1 time each in at least 3 gallons (11.4 liters) of water
and air dry out of direct sunlight.
CAUTION
NOTES
Discoloration of the seal is normal after disinfection.
P
7-23
AGE