1117539, V
. 07
ER
C
E
ODE
RROR
E-30
ERR_MOTOR_SPINUP_F
LUX_HIGH
E-34
ERR_MOTOR_TYPE_UN
KNOWN
E-35
ERR_MOTOR_BLOCKED
_INLET
E-36
ERR_MOTOR_BLOCKED
_OUTLET
E-40
ERR_NVRAM
E-41
ERR_STORAGE_UNIT_R
AM
E-48
ERR_NVRAM_MAX_RET
RIES_EXCEEDED
E-50
ERR_DAILY_VALUES_C
ORRUPT
E-51
ERR_CORRUPT_COMPLI
ANCE_LOG
E
RROR
N
AME
L
EVEL
REBOOT
STOP
CONTINUE
CONTINUE
STOP
REBOOT
REBOOT
CONTINUE
CONTINUE
F
AILED
C
OMPONENT
1. Reseat motor connector,
1. Motor
clear error log and test.
connection
2. Blower Box
2.
If
3. PCA
replace blower and retest.
3. If 2nd retest fails, replace
PCA.
PCA
1. Replace PCA and retest.
1. Air Path
1. Clear air path including
2. PCA
filter and test.
3. Blower Box
2. If test fails, replace PCA
and retest.
3. If retest fails, replace
blower box and retest.
1. Air Path
1. Clear air path including
2. PCA
filter and test.
3. Blower Box
2. If test fails, replace PCA
and retest.
3. If retest fails, replace
blower box and retest.
PCA
1. Replace PCA and test.
PCA
1. If error was NOT Last
Stop Error, clear error log
and test.
2. If error was Last Stop
Error, replace PCA and test.
PCA
1. If error was NOT Last
Stop Error, clear error log
and test.
2. If error was Last Stop
Error, replace PCA and test.