7.1.8 R
EPLACING THE
Kit
Bottom Cover
T
B
O REMOVE THE
OTTOM
1. Turn the Humidifier over.
2. Remove the four # 6 x 1/4" screws securing the Bottom Cover to the Bottom Assembly.
3. Lift the Bottom Cover from the Humidifier.
T
B
O INSTALL THE
OTTOM
1. Align the Bottom Cover onto the Bottom Assembly.
2. Secure the Bottom Cover with the four # 6 x 1/4" screws (torque to 6 IN-LB).
N
: Ensure the wires are routed as shown below before installing the Bottom Cover. Also ensure the wires
OTE
aren't pinched when installing the Bottom Cover.
B
C
OTTOM
OVER
Included in Kit
Bottom Cover
# 6 x 1/4" screws (qty 4)
C
:
OVER
Torque screws to 6 IN-LB when
installing Bottom Cover
F
7-19: B
IGURE
C
:
OVER
F
7-20: W
IGURE
C
OTTOM
OVER
Route wires as shown
R
IRE
OUTING
P
AGE
Tools Required
Torx Screwdriver (T15 is
recommended)
7-13