Prior to testing any DreamStation device, the following must be ensured:
1. The device error log must be cleared.
2. The device real time clock must be accurate.
3. The DreamStation base unit and Humidifier must both be at room temperature. Refer to Section
8.3 for testing environmental specifications.
Refer to Chapter 5 on using the Service Center Tools software to clear the device error log and to reset the real
time clock.
8.3 T
E
ESTING
NVIRONMENT
The DreamStation devices must be tested within the following temperature specifications:
41 º F to 95 º F (+5 º C to 35º C)
8.4 F
T
INAL
ESTING
1. Download the NI LabVIEW 2014 Support Package from my.respironics.com, and install the soft-
ware by accepting all license agreements and default installation locations.
2. Download the DreamStation Device Driver Package from my.respironics.com, and unzip the files
to your PC.
3. Follow the installation instructions within the zip file to install the drivers.
4. Download the DreamStation Field Service Application (FSA) from my.respironics.com, and install
the software by accepting all license agreements and default installation locations. The following
icon will be created on your desktop:
5. Launch the Test Executive from your desktop by double-clicking on the icon.