Clearing The Error And Device Logs; Clearing Therapy Hours And Blower Hours; Setting The Session Id - Respironics DreamStation Service & Technical Reference Manual

Table of Contents

Advertisement

P
5-6
AGE
5.3.2 C
LEARING THE
There should be no errors on the device after repairs are made. If there are any errors logged on
the device that do not affect device functionality, the error(s) must be cleared. Refer to section 5.3
for a list of error codes, descriptions and corrective actions.
The device log cannot be read, however the device log should be cleared on the device as part of
routine servicing.
5.3.3 C
T
LEARING
HERAPY
If the PCA is NOT replaced during device servicing, the therapy hours should only be cleared if the
device is going to a different patient. Otherwise, the therapy hours should remain on the device.
If the PCA is NOT replaced during device servicing, the blower hours should be cleared only if the
Blower is replaced. Otherwise, the blower hours should remain on the device.
5.3.4 S
ETTING THE
The Session ID should be set on the device when the Therapy PCA is
replaced. If the Therapy PCA is not replaced on the device, the Session ID
should not be set.
1. Connect the device to your PC and launch the Service Center Tools Suite.
2. Select QPAP SET SESSIONS ID, then click on the EXECUTE TOOL button.
3. Enter the date manufactured of the PAP device in the DATE MANUFACTURED box, then select
the SET SESSION ID button.
The Manufactured Date is located on the device's serial/model number label in
format YYYY-MM-DD. When entering the date into the software application, do
not include the dashes (-).
E
D
RROR AND
EVICE
H
OURS AND
S
ID
ESSION
L
OGS
B
H
LOWER
OURS
NOTE
NOTE
1117539, V
. 07
ER