Tank; Leaning The; Base - Respironics DreamStation Service & Technical Reference Manual

Table of Contents

Advertisement

1117539, V
. 07
ER
7.2 C

LEANING THE

Hand washing can be performed daily. Dishwashing can be performed once a week.
1. Turn the humidifier setting off, turn the therapy device off, and allow the heater plate and water to
cool.
Allow the humidifier heater plate and water to cool down for approximately 15 minutes
before removing the water tank. A burn may result from: touching the heater plate, coming
in contact with the heated water, or touching the tank pan.
2. Open the humidifier lid with the release lever, and remove the water tank from the humidifier.
3. Wash the water tank in a solution of warm water and a mild liquid dishwashing detergent.
4. Rinse with clean water.
5. Wipe completely on the top and bottom and allow the water tank to air dry.
Use a mild liquid dishwashing detergent only for either hand washing or when
using a dishwasher.
Empty and clean the water tank daily to prevent mold and bacteria growth.
6. Inspect the water tank for damage. If the water tank show signs of wear or damage, replace the

tank.

7.3 C
LEANING THE
Allow the humidifier heater plate and water to cool down for approximately 15 minutes
before removing the water tank. A burn may result from: touching the heater plate,
coming in contact with the heated water, or touching the tank pan.
Before cleaning the humidifier, always remove from the therapy device.
1. Clean the humidifier base and heater plate by wiping it with a damp cloth. Allow the platform to air
dry before reconnecting to the therapy device.
2. Inspect the humidifier base for any damage and replace it if necessary.
3. Clean the humidifier outlet port by using a damp bottle brush or a damp cloth. Insert the brush or
cloth approximately 2.75 inches (7 cm) into the outlet opening while cleaning.
W
T
ATER
ANK
WARNING
CAUTION
WARNING
H
B
UMIDIFIER
ASE
WARNINGS
P
7-21
AGE