• Your mobile device has Bluetooth Secure Simple Pairing (SSP)
The following icon will pop-up on your therapy device screen with a 6 digit number and "Pair?":
This number is a six digit passkey generated during SSP. Verify that the six digit SSP passkey is the same on
both the mobile device and therapy device. Rotate the Control Dial between "yes" or "no", and then press the
Control Dial to choose. If "no" is selected, or the pop-up screen times out after 30 seconds, the device will
reject the pair request. If "yes" is selected, the therapy device will acknowledge the six digit SSP passkey. If the
mobile device also acknowledges the request, the two will now be paired and ready to connect using
DreamMapper.
• Your Bluetooth enabled mobile device does not support Bluetooth SSP
Your mobile device will prompt you to enter a pin code. Enter "1008" on your mobile device. The following icon
will pop-up on your therapy device screen with the number "001008" and "Pair?":
Rotate the Control Dial between "yes" or "no", and then press the Control Dial to choose. If "no" is selected, or
the pop-up screen times out after 30 seconds, the device will reject the pair request. If "yes" is selected, the
therapy device will acknowledge the 001008 passkey. If the mobile device also acknowledges the request, the
two will now be paired and ready to connect using DreamMapper.
Note: Do NOT select "yes" on the pop-up screen unless you are currently trying to pair your devices. This will
ensure that only your mobile device connects to your therapy device.
4.13 A
CCESSORIES
There are several accessories available for the DreamStation system such as a Humidifier, Cellular Modem,
Wi-Fi Accessory or a Link Module. When using optional accessories, always follow the instructions enclosed
with the accessories.
Pins of connectors should not be touched. Connections should not be made to these
connectors unless ESD precautionary procedures are used. Precautionary procedures include
methods to prevent build-up of electrostatic charge (e.g., air conditioning, humidification,
conductive floor coverings, non-synthetic clothing), discharging one's body to the frame of the
equipment or system or to earth or a large metal object, and bonding oneself by means of a
wrist strap to the equipment or system or to earth.