Status Byte
0 1 X 0 0 1 1 1
0 1 X 0 1 0 0 0
0 1 X 0 1 0 0 1
0 1 X 0 1 0 1 0
0 1 X 0 1 0 1 1
0 1 X 0 1 1 0 0
0 1 X 0 1 1 1 0
0 1 X 0 1 1 1 1
Cavro® XLP 6000 Modular Syringe Pump Operating Manual, 734237-C, en
Hex # if Bit 5 =
47h
67h
48h
68h
49h
69h
4Ah
6Ah
4Bh
6Bh
4Ch
6Ch
4Eh
6Eh
4Fh
6Fh
Error Reporting Examples
[A7000R]
Does not move the plunger and reports a "No Error" status;
when queried ([Q] command), returns error. A
second try returns error 3 (67).
[P6000P600R]
Moves to position 6000, then stops. A [Q] command returns
an error.
[t2000R]
Returns an invalid command error immediately. The pump
status is "Not Busy".
[A6000t2000R]
Returns an invalid command error immediately. The pump is
"Not Busy."
Valve in Bypass
Returns an error immediately; when queried ([Q] command),
[A1000R]
does not return an error.
Dec # if Bit 5 =
71
103
72
104
73
105
74
106
10
75
107
11
76
108
12
78
110
14
79
111
15
3 - Software Communication
Error Codes and Pump Status
Error Code
7
Device not Initialized
8
Internal failure
9
Plunger Overload
Valve Overload
Plunger Move Not Allowed
Internal Failure
A/D converter failure
Command Overflow
3 - 51