Figure 2-6 Xlp 6000 Valve Installation (3-Port Valve Shown) - tecan cavro xlp 6000 Operating Manual

Modular syringe pump
Table of Contents

Advertisement

2 - Hardware Setup
Installing Components
Figure 2-6
2.5.2
2 - 12
XLP 6000 Valve Installation (3-Port Valve Shown)
Installing a Syringe
To install a syringe, follow these steps:
1
Remove the plunger lock screw.
2
Install the syringe as shown in Figure 2-7, following these steps:
a. Screw the syringe into the valve.
b. Pull the syringe plunger down to the plunger holder assembly.
c. Align the plunger button through hole to the carriage mounting hole.
d. Slide the plunger lock screw through the plunger button and fasten to the
carriage.
Note: Make sure the plunger lock screw is securely tightened and the plunger
button is free to move on the plunger lock screw.
Cavro® XLP 6000 Modular Syringe Pump Operating Manual, 734237-C, en