Switching Power Supplies
Be sure to check carefully the minimum load requirement of the power supply.
Typically, switching supplies have a minimum load requirement of up to 10% of
the rated output current.
Note: The XLP 6000 idle current is less than 10% of the full running current.
For example, in a system with multiple XLP 6000 pumps, a 24V 5-amp switcher
with a minimum load less that 500mA may not provide sufficient current when the
XLP 6000 motors are idle and all other devices are in a low current state. If the
XLP 6000 is the only load on the 24V supply, a switcher should have a minimum
load specification of 50mA or less. An appropriate external power resistor can be
used to ensure that the minimum load is met.

Cabling

A single cable supplies both power and communications to each XLP 6000. A
unique address identifies each pump module. For more information, see "Address
Switch Settings" later in this chapter. See also Chapter 3, "Software
Communication"."
Power requirements are described in Section 2.1, Power and Electrical
Considerations.
DA-15 Connector Pin Assignments
Pin
Function
1
24V DC
2
RS-232 TxD line
3
RS-232 RxD line
4
Reserved
5
CAN high signal line
6
CAN low signal line
7
Auxiliary input #1
8
Auxiliary input #2
9
Ground
10
Ground
11
RS-485 A line
2 - Hardware Setup
Remarks
Output data
Input data
TTL level
TTL level
Power and logic
Power and logic
Data +
Cabling
2 - 3