tecan cavro xlp 6000 Operating Manual page 58

Modular syringe pump
Table of Contents

Advertisement

3 - Software Communication
Initialization
3 - 26
<n
>
0
2
1..X
<n
>
0
3
1..X
W<n
>
Initialize Plunger Drive
1
The [W] command initializes the plunger drive only (commonly used for
valveless pumps). Because the valve is not initialized, only plunger force and/
or speed can be set.The default initialization speed is 500 pulses per second.
n
= Set initialization plunger force/speed
1
The parameters are described below.
W Parameter
Value
<n
>
0
1
1
2
10-40
Cavro® XLP 6000 Modular Syringe Pump Operating Manual, 734237-C, en
Sets initialization input port to port 1 (default)
Sets initialization input port for distribution valves,
where X is the number of ports on the valve.
Sets initialization output port to port X (default), where X
is the number of ports on the valve.
Sets initialization output port for distribution valves,
where X is the number of ports on the valve.
Description
Initializes at full plunger force and at default initialization
speed (default)
Initializes at half plunger force and at default initializa-
tion speed
Initializes at one-third plunger force and at default ini-
tialization speed
Initializes at full force and at speed code <n
command <S> for a list of speed codes.
>. See