Error Codes And Status Byte - tecan cavro xlp 6000 Operating Manual

Modular syringe pump
Table of Contents

Advertisement

G - Command Quick Reference

Error Codes and Status Byte

G.10 Error Codes and Status Byte
Status Byte
7 6 5 4 3 2 1 0
0 1 X 0 0 0 0 0
0 1 X 0 0 0 0 1
0 1 X 0 0 0 1 0
0 1 X 0 0 0 1 1
0 1 X 0 0 1 1 0
0 1 X 0 0 1 1 1
0 1 X 0 1 0 0 1
0 1 X 0 1 0 1 0
0 1 X 0 1 0 1 1
0 1 X 0 1 1 0 0
0 1 X 0 1 1 1 0
0 1 X 0 1 1 1 1
G.11 DA-15 Connector Pin Assignments
Pin
1
2
3
4
5
6
7
8
9
G - 8
Hex # if Bit 5 =
0
or 1
40h
60h
41h
61h
42h
62h
43h
63h
46h
66h
47h
67h
49h
69h
4Ah
6Ah
4Bh
6Bh
4Ch
6Ch
4Eh
6Eh
4Fh
6Fh
Function
24 V DC
RS-232 TxD line
RS-232 RxD line
Unused
CAN high signal line
CAN low signal line
Auxiliary input #1
Auxiliary input #2
Ground
Cavro® XLP 6000 Modular Syringe Pump Operating Manual, 734237-C, en
Dec # if Bit 5 =
Error Code
0
or 1
Number
64
96
0
65
97
1
66
98
2
67
99
3
70
102
6
71
103
7
73
105
9
74
106
10
75
107
11
76
108
12
78
110
14
79
111
15
Remarks
Output data
Input data
TTL level
TTL level
Power and logic
Description
No error
Initialization
Invalid command
Invalid operand
EEPROM failure
Device not initialized
Plunger overload
Valve overload
Plunger move not
allowed
Internal failure
A/D converter failure
Command overflow

Advertisement

Table of Contents
loading

Table of Contents