Initialization; Initialization - tecan cavro xlp 6000 Operating Manual

Modular syringe pump
Table of Contents

Advertisement

3.4
3.4.1
Table 3-6
Cavro® XLP 6000 Modular Syringe Pump Operating Manual, 734237-C, en
Table 3-5
Write Pump Configuration Command Values
Value<n>
47
51
52
53
54
57
Note: [U] commands take effect upon the pump's next power-up.

Initialization

Initialization Forces
IInitialization moves the plunger to the top of travel, then backs off a user-
specified number of increments (see k command) and sets this as position 0.
Also, the input and output positions of the valve are assigned depending on the
initialization command. All other command parameters are reset to default values.
The top of the syringe is recognized when upward movement of the plunger
causes an overload condition.
The force at which the plunger presses against the top of the syringe can be
controlled via a parameter after the Initialization command (possible values are 0,
1 and 2).
Table 3-6 lists the recommended initialization force for each type of syringe.
Caution! To retain the integrity of the seal on smaller syringes, use a lower
initialization force than that for larger syringes. The default initialization speed is
500 Hz.
Recommended Initialization Forces by Syringe
Syringes
1.0 mL and larger
250, 500 µL
50, 100 µL
3 - Software Communication
Description
Set RS-232/RS-485 Baud rate to 38400
Set CAN Bus Baud rate to 100K
Set CAN Bus Baud rate to 250K
Set CAN Bus Baud rate to 500K
Set CAN Bus Baud rate to 1M
Set CAN Bus Baud rate to 125K
Force
Full
Half
Third

Initialization

3 - 23