v <n>
Set Start Speed
The [v] command sets the speed at which the plunger begins its movement, in
pulses/sec. The plunger will then ramp up (slope) to the top speed. The start
speed should always be less than the top speed.
<n>
Parameter
Command
Value
v
50-1000
V <n>
Set Top Speed
The [V] command sets the top speed in pulses/second. This command may
be sent while a command string is already executing. (See section on
Changing Speed on the Fly, earlier in this chapter.)
<n>
Parameter
Command
Value
V
5-6000
Note: Syringes 2.5 mL and larger may require slower speeds. Users must
determine the appropriate speeds for their applications.
Cavro® XLP 6000 Modular Syringe Pump Operating Manual, 734237-C, en
8
20000
9
22500
10
25000
11
27500
12
30000
13
32500
14
35000
15
37500
16
40000
17
42500
18
45000
19
47500
20
50000
Default
Value
Description
900
Set start speed in pulses/sec
Default
Value
Description
900
Set top speed in pulses/sec