tecan cavro xlp 6000 Operating Manual page 34

Modular syringe pump
Table of Contents

Advertisement

3 - Software Communication
XLP 6000 Addressing Scheme
Table 3-2
Single Device
Switch
Hex
Setting
Address
0
31
1
32
2
33
3
34
4
35
5
36
6
37
7
38
8
39
9
3A
A
3B
B
3C
C
3D
D
3E
E
3F
F
Self Test
3 - 2
Address Switch Settings in Hex (ASCII)
Dual Device
ASCII
Hex
Address
Address
1
41
2
3
43
4
5
45
6
7
47
8
9
49
:
;
4B
<
=
4D
>
?
4F
The user can communicate with all pumps in the chain by using address "5Fh," for
example to initialize all pumps at once. Then each pump can be controlled
independently by using addresses "31h" to "3Fh."
Note: Multiple address commands cannot be used to determine device status or
to request reports. Each device must be queried separately to gather status or
generate a report.
Cavro® XLP 6000 Modular Syringe Pump Operating Manual, 734237-C, en
Quad Device
ASCII
Hex
Address
Address
A
51
C
E
55
G
I
59
K
M
5D
O
All Devices
ASCII
Address
Address
Q
5F
U
Y
]
Value to
Send