tecan cavro xlp 6000 Operating Manual page 39

Modular syringe pump
Table of Contents

Advertisement

3.2.2
Table 3-4
Parameter
Baud rate
Data bits
Parity
Stop bit
1
2
2+n
3+n
1
2
3
Cavro® XLP 6000 Modular Syringe Pump Operating Manual, 734237-C, en
Master Address
The master address is the address of the host system. This should always be
30h (ASCII value "0").
Status and Error Codes
The status and error codes define pump status and signal error conditions.
For a description of status and error codes, see "Error Codes and Pump
Status" in this chapter.
Data Terminal (DT) Protocol
The DT protocol can be used easily from any terminal or terminal emulator
capable of generating ASCII characters at 9600 baud, 8 bits, and no parity.
DT Protocol
Setting
9600 or 38400
8
None
1
(see "DT Protocol Command Block Characters" for details)
Start command (ASCII "/" or 2Fh)
Pump address
Data block (length n)
Carriage Return ([CR] or 0Dh)
(see "DT Protocol Answer Block Characters" for details)
Start answer (ASCII "/" or 2Fh)
Master address (ASCII "0" or 30h)
Status character
Character Format
Command Block
Answer Block
3 - Software Communication
Communication Protocols
3 - 7

Advertisement

Table of Contents
loading