B - Plunger Information
Plunger Time Calculations
B.2.2 Move Calculations
B - 2
Symbol
Name
t
Total Move Time
A
Ramp Up increments
1
A
Ramp Up increments
2
A
Ramp Up increments
3
Note:
•
Cutoff speed cannot be less than start speed.
•
During aspiration, v is used in place of c.
•
The XLP 6000 uses 6000 full increments per stroke.
Case 1. Start, Top, and Cutoff Speeds are Equal or Top Speed is Less than
50 Hz
Case 1 is used when v = V = c or V < 50
Diagram of move:
Speed
[Hz]
V=c=v
Calculation:
v = 900 Hz
V = 900 Hz
c = 900 Hz
Total Move Time:
∗
∗
2
A
2 3000
=
=
t
V
900
Cavro® XLP 6000 Modular Syringe Pump Operating Manual, 734237-C, en
Range (n)
t
+t
+t
1
2
3
t
time in s
L = 14
A = 3000 full increments
=
seconds
6 67
.
Unit
Seconds
Half increments
Half increments
Half increments