Plunger Time Calculations - tecan cavro xlp 6000 Operating Manual

Modular syringe pump
Table of Contents

Advertisement

B - Plunger Information

Plunger Time Calculations

B.2.2 Move Calculations
B - 2
Symbol
Name
t
Total Move Time
A
Ramp Up increments
1
A
Ramp Up increments
2
A
Ramp Up increments
3
Note:
Cutoff speed cannot be less than start speed.
During aspiration, v is used in place of c.
The XLP 6000 uses 6000 full increments per stroke.
Case 1. Start, Top, and Cutoff Speeds are Equal or Top Speed is Less than
50 Hz
Case 1 is used when v = V = c or V < 50
Diagram of move:
Speed
[Hz]
V=c=v
Calculation:
v = 900 Hz
V = 900 Hz
c = 900 Hz
Total Move Time:
2
A
2 3000
=
=
t
V
900
Cavro® XLP 6000 Modular Syringe Pump Operating Manual, 734237-C, en
Range (n)
t
+t
+t
1
2
3
t
time in s
L = 14
A = 3000 full increments
=
seconds
6 67
.
Unit
Seconds
Half increments
Half increments
Half increments

Advertisement

Table of Contents
loading

Table of Contents