2680A/2686A
Users Manual
Displaying the Totalizer Status
Perform the procedure in Table 2-7 to display the instrument totalizer status. The
instrument updates the totalizer display once per second. (See Figure 2-13 for
examples.) To initialize the totalizer count, cycle the instrument power. You can
also configure Fluke DAQ software to initialize the totalizer count when it starts
logging.
•
Maximum Totalizer Count The totalizer counts switch closures or voltage
transitions with a maximum count of 4,294,967,295.
•
Totalizer Status Display The instrument displays the totalizer status in two
five digit segments; high (HI) and low (LO). Using the maximum count of
4,294,967,295 as an example, the HI five digit segment would display 42949
and the LO five digit segment 67295.
G
C
B
G
2-30
Table 2-7. Displaying Totalizer Status
Press the DIO (Digital I/O) key to view the totalizer status. You can view the
totalizer status whether the instrument is scanning or not scanning.
Press an up/down arrow key to advance to the totalizer display. The 10 digits
display in a five digit tot:HI count and five digit tot:LO count.
Press the DIO key again to exit. Fluke DAQ software also provides a display of
the totalizer status.