Displaying the Totalizer Status

Perform the procedure in Table 2-7 to display the instrument totalizer status. The
instrument updates the totalizer display once per second. (See Figure 2-13 for
examples.) To initialize the totalizer count, cycle the instrument power. You can
also configure Fluke DAQ software to initialize the totalizer count when it starts
logging.
Maximum Totalizer Count The totalizer counts switch closures or voltage
transitions with a maximum count of 4,294,967,295.
Totalizer Status Display The instrument displays the totalizer status in two
five digit segments; high (HI) and low (LO). Using the maximum count of
4,294,967,295 as an example, the HI five digit segment would display 42949
and the LO five digit segment 67295.
G
C
B
G
2-30
Table 2-7. Displaying Totalizer Status
Press the DIO (Digital I/O) key to view the totalizer status. You can view the
totalizer status whether the instrument is scanning or not scanning.
Press an up/down arrow key to advance to the totalizer display. The 10 digits
display in a five digit tot:HI count and five digit tot:LO count.
Press the DIO key again to exit. Fluke DAQ software also provides a display of
the totalizer status.