Fluke 2680A User Manual page 168

Data acquisition system/data logging system
Hide thumbs Also See for 2680A:
Table of Contents

Advertisement

2680A/2686A
Users Manual
2. Connect the 5700A to Channel 1 of the First Module When testing each
module, connect a cable from the Output VA HI and LO connectors of the
5700A to the Universal Input module terminals for channel 1 be certain to
connect the 5700A HI to terminal H and LO to terminal L. Insert the Universal
Input module into the instrument under test (Figure 4-3).
Device Type
Multifunction
Calibrator
Digital Multimeter
Signal Generator
Oscilloscope
Multiproduct Calibrator
Function/Signal Generator
Decade Resistance Source
4-10
Table 4-2. Recommended Test Equipment
Minimum Specifications
Range: 90 mV to 300 V dc
Accuracy: 0.002%
Frequency
1 kHz
100 kHz
10 kHz
Ohms
190Ω
1.9 kΩ
19 kΩ
190 kΩ
1.9 MΩ
General Purpose Measurement
Sine wave. 0.5 to 1 V rms, 10 Hz to 5 kHz
Alternative Equipment List
Device Type
DC Voltage:
AC Voltage:
Voltage
Accuracy
29 mV to 300 V
0.05%
15 mV to 300 V
0.5%
Frequency:
1 V rms
0.01%
Ohms:
Accuracy
0.005%
0.005%
0.005%
0.005%
0.005%
Recommended Model
Fluke 5520
Philips PM5139 or HP33120A or equivalent
General Resistance RDS 66A or equivalent
Recommended Model
Fluke 5700A
Fluke 77
Philips PM5193

Advertisement

Table of Contents
loading