the network number), the data is sent through the gateway to reach the host
PC.
The network number of the instrument must match that of the gateway. For
example, if the gateway IP address is 129.196.180.93, and the subnet mask is
255.255.255.0, the network number is 129.196.180.0, and the instrument IP
address must be in the range 129.196.180.1 to 129.196.180.254.
Table 2-14. Reviewing and Setting the Subnet Mask and Default Gateway
F
Press the COMM key to review the parameters, or press and hold the COMM
key for 3 seconds to set the parameters.
C
Press the up/down arrow keys until dgAtE (default gateway) appears in the
B
primary display (COMM appears in the secondary display).
I
Press the ENTER key. dgAtE appears in the secondary display, and ON or
OFF is in the primary display.
C
Press the up/down arrow keys to select either ON or OFF when in set mode.
B
I
Press the ENTER key to make the selection (pressing any other key will cancel
set operations.). If you select ON, the subnet mask appears. The subnet mask
display consists of four 3 digit segments: Sub0. Sub1. Sub2. Sub3.
Press the left/right arrow keys to select the desired number in each segment.
D
The selected digit is highlighted and the segment, for example Sub:0, appears
E
in the secondary display.
C
Press the up/down arrow keys to select the desired number 0 to 9, for the
positioned subnet mask digit. In this manner, select all 12 subnet mask digits.
B
I
Press the ENTER key to make the selection (pressing any other key will cancel
set operations). If you enter an incorrect subnet mask, Error appears for 2
seconds, and the subnet mask selection stays displayed. Otherwise, the
default gateway IP address appears. The default gateway display consists of
four 3 digit segments: gAt0.gAt1.gAt2.gAt3.
D
Press the left/right arrow keys to select the desired number in each segment.
The selected digit becomes highlighted and the segment, e.g. gAt:0, appears in
E
the secondary display.
C
Press the up/down arrow keys to select the desired number 0 to 9, for the
positioned default gateway IP digit. In this manner, select all 12 default
B
gateway IP digits.
I
Press the ENTER key to make the selection (pressing any other key will cancel
set operations). If you enter an incorrect default gateway IP, Error displays for 2
seconds, and the default gateway IP selection stays displayed. Otherwise, the
procedure exits.