Fluke 2680A User Manual page 238

Data acquisition system/data logging system
Hide thumbs Also See for 2680A:
Table of Contents

Advertisement

2680A/2686A
Users Manual
Custom-385 RTD
The Custom-385 RTD linearization solves the Callendar Van-Dusen equation for
positive temperatures and uses a polynomial approximation to the Callendar Van-
Dusen equation for negative temperatures. In addition to R
specified using the Fluke DAQ software. δ and β are fixed. These fixed
coefficients are the modified DIN 43760 coefficients that conform to ITS90:
δ = 1.49978574
β = 0.10863383
Figure D-1 shows the accuracy envelope for a 385 RTD that has δ and β that
match the above coefficients. Because of the error introduced by the linearization
at low temperatures, it is recommended that use of the Custom-385 RTD type be
limited to temperatures between -100 and 600 °C.
1.5
1.0
0.5
0.0
-0.5
-1.0
-1.5
D-2
Accuracy Envelope: Custom RTD, 385 Probe
1 Year Slow: 18-28 Degrees C
Alpha: 0.00385055 Delta: 1.49978574 Beta: 0.10863383
FAI
PAI
-100
0
100
Temperature - Degrees C
Figure D-1. 385 RTD
200
300
400
, α may also be
0
500
600
alg61f.eps