The following table provides additional safety information.
Specification
Size
Weight
Power
EMC
Safety
Operating Temperature
Range
Storage Temperature
Range
Relative Humidity
Altitude
General Specifications
473 mm (18.6 in) x 432 mm (17 in) x 237 mm (9.3 in)
2680A/2686A (empty)
2680A – FAI
2680A – PAI
2680A – DIO
100 – 240 V ac (no switching required), 50 to 60 Hz, 100 VA
maximum or optional 9 V dc to 45 V dc, 35 W maximum
EN50082-2
EN55022-1
EN55011 class A
EN610000-4-2,3,4,6,8
EN61326
EN61010-1, CAT II
CSA C22.2 No. 1010.1
o
o
o
-20
C to 60
C (-4
F to +140
o
o
o
-40
C to 70
C (-40
F to +158
90% maximum for -10
75% maximum for 28
50% maximum for 35
(3 MΩ range, reduce humidity rating by 25% for 1 hour warm-up.
3 MΩ range meets full humidity ratings with 2 hour warm-up)
Operating: 2,000 m (6,562 ft) maximum
Non-operating: 12,200 m (40,000 ft) maximum
Characteristic
8.6 Kg (18.9 lb)
0.8 Kg (1.8 lb)
1.2 Kg (2.7 lb)
0.8 Kg (1.8 lb)
o
F)
o
F)
o
o
o
C to 28
C (14
F to +82
o
o
o
C to 35
C (82
F to +95
o
o
o
C to 60
C (95
F to +140
o
F)
o
F)
o
F)