2680A/2686A
Users Manual
For the NetDAQ 2640A and 2645A, the scan trigger, intervals,
reading rate, and drift correction are set on the Instrument
Configuration dialog instead of the Module settings dialog.
Conditional scan storage is not available on the NetDAQ
instruments.
If you are configuring a DIO module, the Totalizer Configuration dialog appears
instead of the Module settings dialog.
The dialog box entries include:
Start Count
Used to specify a start count.
Debounce
Select debounce to ignore signals with a period less than 2 ms (500 Hz);
otherwise, signals up to 5 kHz can be counted.
Direction
If you entered a start count, the start count is reloaded when the totalizer
reaches the terminal count. The totalizer can count up or down to zero.
You can also highlight the IO module and click the
settings including digital IO points and relays.
3-22
Note
alg111s.bmp
button for more DIO